Roche/Hitachi cobas c 311 analyzer
Roche Diagnostics February 2017
12 Host Interface Manual – Version 1.8
function is enabled in the Auto Rerun mode during analysis, an
inquiry is made for the Auto Rerun selections. When this function is
not enabled, no inquiry for the Auto Rerun Selection is made.
Auto Rerun TS Timeout
Check box
On/Off and input 10-9999
This function applies to the reception of Auto Rerun Test Selection
Information in Realtime Communications. The range is 10-9999
seconds, with a default of 18 seconds (depending on TS Priority
Mode). If the function is not enabled, the analyzer waits indefinitely
for Rerun Test Selection Information.
Manual Rerun T S
Check box On/Off
This function applies to the Test Selection Information inquiry for
Manual Rerun samples in Realtime Communication. When this
function is enabled, an inquiry is made for the test selections for
rerun samples during analysis. Any samples that have previously
been processed create a TS inquiry to the host. When this function is
not enabled, no TS inquiry for Manual Rerun samples is made.
Stat TS
Check box On/Off
This function applies to the Test Selection Information inquiry for
Stat samples in Realtime Communications. When this function is
enabled, an inquiry is made for the test selections for Stat samples
during analysis. When this function is not enabled, no TS inquiry for
Stat samples is made.
TS Inquire Always
Check box
On/Off
This function applies to the Test Selection Information inquiry for
routine and Stat samples in Realtime Communications. This does not
apply to TS information inquiry for rerun samples, Auto Rerun and
Manual Rerun. When this function is enabled, an inquiry is made
whether the TS information is in the analyzer or not. When not
enabled, an inquiry is made only for the samples that do not have test
selections.
TS Ask in Barcode Read Error
Check box On/Off
If this function is enabled, and the analyzer cannot read a sample
barcode, “***...” (13 or 22 characters of ‘*’) is sent as the sample ID,
and the rack ID and position are sent to the host as key information
to identify the sample. If the host has the table of the rack ID, position
and Sample ID, the host can send back to the analyzer the correct
Sample ID instead of ‘***...’
Because Host protocol and text field is same as cobas6000, Rack ID field is used
in cobas c 311 analyzer communication text.
The only currently known use for this function is with a Sample
Handling System that ‘manufactures’ sample aliquots for the analyzer.
This type of system would track samples by position on S.Disk.
Unless the Sample Handling System can perform this function, it is
strongly recommended that it not be used