Thank you very much for using SUNX products. Please read this
Instruction Manual carefully and thoroughly for the correct and
optimum use of this product. Kindly keep this manual in a con-
venient place for quick reference.
SPECIFICATIONS
1
Indicators
12 to 24V DC % Ripple P-P 10% or less
+10
-
15
Item
Type of pressure
Model No. DP-M2 DP-M2A
Type Standard With analog current output
50mA or less 75mA or less
Differential pressure
Rated pressure range 0 to 2.00 kPa.D {0 to 204 mmH
2O.D}
Hysteresis 0.01 kPa.D {1mm H2O.D}
Repeatability Within r1% F.S.
Displayable
pressure range
-0.05 to 2.10 kPa.D {-5 to 210 mmH
2O.D}
Operation Orange LED (lights up when the comparative output is ON)
Pressure unit
Red LED (The indicator corresponding to the selected unit
lights up during the sensing mode.)
M1 setting Red LED (blinks in the M1 setting mode)
M2 setting Red LED (blinks in the M2 setting mode)
Response time 10ms or less
Short-circuit protection
Incorporated
Set pressure range 0 to 2.00 kPa.D {0 to 204 mmH2O.D}
Setting resolution 0.01 kPa.D {1 mmH
2O.D}
Pressure withstandability 6 kPa.D {612 mmH
2O.D}
Applicable fluid Non-corrosive gas
Display
3 digit red LED display (Sampling rate: 4 times/sec. approx.)
Comparative output
NPN open-collector transistor
Maximum sink current: 100mA
Applied voltage: 30V DC or less
(between comparative output and 0V)
Residual voltage: 1V or less (at 100mA sink current)
0.4V or less (at 16mA sink current)
Output operation
Supply voltage
Ambient temperature
0to+50 (No dew condensation), Storage: -10 to +60
Ambient humidity 35 to 85% RH, Storage: 35 to 85% RH
Pressure port Ǿ4.8mm resin pipe
Weight 75g approx.
Temperature characteristics
Over ambient temperature range 0 to +50:
within r3% F.S. of detected pressure at 25
Material
Front case: ABS, Rear case: ABS
LED display: Acrylic, Pressure port: PA
Cable
0.18mm
2
3-core oil resistant
cabtyre cable, 2m long
0.18mm
2
4-core oil resistant
cabtyre cable, 2m long
Current consumption
㧙
Analog current output
Selectable either Normally open (NO) or Normally closed
(NC) by the keys.
Refer to '˴˴SETTING'[Output mode and output op-
eration] for more details.
8
Output current: 4 to 20mA
0to1.96kPa.D
{0 to 200 mmH
2O.D}
Zero-point: Within 4mA r1% F.S.
Span: Within 16mA r3% F.S.
Linearity: Within r1% F.S.
Load resistance: 0 to 250ǡ
CAUTIONS
2
٨
٨
٨
٨
٨
٨
٨
٨
٨
٨
٨
٨
٨
٨
Use within the rated pressure range.
Make sure that the power supply is off while wiring.
Take care that wrong wiring will damage the sensor.
Verify that the supply voltage variation is within the rating.
If power is supplied from a commercial switching regulator, en-
sure that the frame ground (F.G.) terminal of the power supply is
connected to an actual ground.
In case noise generating equipment (switching regulator, inverter
motor, etc.) is used in the vicinity of this sensor, connect the
frame ground (F.G.) terminal of the equipment to an actual
ground.
Do not use during the initial transient time (0.5 sec.) after the
power supply is switched on.
Do not run the wires together with high-voltage lines or power
lines or put them in the same raceway. This can cause malfunc-
tion due to induction.
Avoid dust, dirt, and steam.
Take care that the sensor does not come in direct contact with
water, oil, grease, organic solvents, such as, thinner etc., or
strong acid, and alkaline.
Do not insert wires, etc, into the pressure port. The diaphragm
(pressure sensing device) will get damaged and correct opera-
tion shall not be maintained.
Do not operate the keys with pointed or sharp objects.
Make sure that stress by forcible bend or pulling is not applied
directly to the sensor cable joint.
This is a CE conformity product complying with EMC Directive.
The standard with regard to immunity that applies to this product
is EN 61000-6-2, and in order to meet the standard, every cable
connected to this product must be within 10m with 0.3mm
2
,or
more, cable. However, in case CE conformity is not required, the
cable length can be up to 100m with 0.3mm
2
, or more, cable.
DP-M series is designed for use with non-corrosive gas. It cannot
be used for liquid or corrosive gas.
INSTRUCTION MANUAL
MJE-DPM No.0561-04
DP-M Series
LED Display 㨯 Digital Differential Pressure Sensor
Pressure Sensor
For use outside Japan
PIPING
3
Recommended tube
Notes: 1)
2)
3)
TYGON is registered trademarks of SAINT-GOBAIN
.
Ensure that excessive pressure is not applied to the pressure ports. Since this
sensor is designed for detecting small pressures, if excessive pressure or shock
is applied to the pressure ports, the diaphragm (pressure sensing device) in the
sensor may get damaged.
Please do not compress the tube. If the tube is compressed, pressure exceeding
the rated value may be generated, damaging the diaphragm (pressure sensing
device).
Pressure Hi-port
Fix tubes on ports securely
Pressure Lo-port
Flexible tubes
(silicone tube etc.)
TYGON tube, size: internal dia 4mm, external dia 6mm, made
by SAINT-GOBAIN.
Not good
٨
٨
Apply higher pressure to the Hi-port and lower pressure to the
Lo-port.
Use flexible tubes (silicone tube etc.) that can fit the pressure
ports, Ǿ4.8mm in diameter. The tubes should cover more than
half the length of the pressure ports.
٨
٨
٨
Never use this product as a sensing device for per-
sonnel protection.
In case of using sensing devices for personnel pro-
tection, use products which meet laws and stand-
ards, such as OSHA, ANSI or IEC etc., for person-
nel protection applicable in each region or country.
In case this sensor is used within Japan, SI unit
must be used since use of pressure units in Japan
is restricted to SI units.
WARNING